Lumasense-technologies IS 320 User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 34
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
IS 320 IGA 320 Manual General 10
Repeatability:
( =1, t
90
=1 s,
T
amb.
=23°C)
0.2% of measured value in °C + 1 °C
Protection
C
lass:
IP65 (IEC 60529)
Ambient
Temperature:
0 to 70 °C
Storage
Temperature:
-20 to 70 °C
Rel.
H
umidity:
No
n
condensing conditions
Weight:
0.3 kg
Housing:
Stainless steel
Connector:
8
pin
connector
Dimensions
[mm]:
145.5
54
103
Ø25
Ø23.5
M
2
5
x
1
.
5
36,5
37
Sighting:
Built
-
in LED targeting light
CE
-
L
abel:
According to EU directives about electromagnetic immunity
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4 Accessories (option)
A wide variety of accessories guarantee easy installation of the pyrometers. The following
overview shows a selection of suitable accessories (see also Chapter 9, Reference Numbers):
Mounting
A mounting angle accessory is available for mounting
and aligning the pyrometer to the measured object.
This enables an easy and fast adjustment of the
pyrometer.
Displays
Several digital displays are available for temperature
indication.
Air Purge
The air purge protects the lens from contamination
by dust and moisture. It has to be supplied with dry
and oil-free pressurized air and generates an air
stream shaped like a cone.
90° Mirror
The 90° mirror enables the capture of objects at an
angle of 90° to the pyrometer axis.
Mounting Angle
Digital Display
Air Purge
90°
-
Mirror
Page view 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 33 34