Page 1 - OPERAT I ON MANUAL

OPERAT I ON MANUALIMPAC PyrometerIS 320  IGA 320

Page 2

IS 320 • IGA 320 Manual General  10Repeatability:( =1, t90=1 s,Tamb.=23°C)0.2% of measured value in °C + 1 °CProtectionClass:IP65 (IEC 60529)Ambien

Page 3 - Contents

IS 320 • IGA 320 Manual General  113 Controls and Installation3.1 Electrical InstallationThe pyrometers are powered by a voltage of 24 V DC (possibl

Page 4

IS 320 • IGA 320 Manual General  12Connecting to the RS485 Interface/BaudrateThe pyrometer is operated in half-duplex mode. To prevent reflectionsdu

Page 5 - 1 General

IS 320 • IGA 320 Manual General  13becomes diffuse. The targeting light only marks the center of the measuring spot, not the size.A measuring tape c

Page 6 - 1.4 Unpacking the Instrument

IS 320 • IGA 320 Manual General  14The spot sizes in the table abovewill only be achieved at theadjusted distances. Decreasing orincreasing the meas

Page 7

IS 320 • IGA 320 Manual General  154 Settings/Parameter DescriptionsThe pyrometer is equipped with a wide range of settings to ensure accurate tempe

Page 8

IS 320 • IGA 320 Manual General  16same temperature. Different materials have different emissivities, ranging from 0% to 100%.The emissivity of an o

Page 9 - 2 Introduction

IS 320 • IGA 320 Manual General  174.9 Clear Time of the Maximum/Minimum Value Storage (tCL)If maximum value storage is on, the highest last tempera

Page 10 - 2.4 Accessories (option)

IS 320 • IGA 320 Manual General  184.9.1 Max/MinWhen entering a reset time interval other than 0, it must also be determined with “max” or“min” if t

Page 11 - 3 Controls and Installation

IS 320 • IGA 320 Manual General  19 Close above: The contact closes (switches to ground), if the temperature entered in the“SP1“ box is exceeded. I

Page 12 - 3.2 Mechanical Installation

Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us

Page 13 - 3.3 Optics

IS 320 • IGA 320 Manual General  20 To ensure consistent document formatting, this page was intentionally left blank.

Page 14

IS 320 • IGA 320 Manual Software InfraWin  215 Software InfraWinThe operating and analyzing InfraWin software is included with delivery of the pyrom

Page 15 - 4.5 Emissivity 

IS 320 • IGA 320 Manual Software InfraWin  22To ensure consistent document formatting, this page was intentionally left blank.

Page 16 - 4.7 EmiAutoFind

IS 320 • IGA 320 Manual Maintenance  236 Maintenance6.1 SafetyAttention during service to the pyrometer: If the pyrometer is integrated into arunnin

Page 17

IS 320 • IGA 320 Manual Maintenance  24To ensure consistent document formatting, this page was intentionally left blank.

Page 18 - 4.9.1 Max/Min

IS 320 • IGA 320 Manual Troubleshooting  257 TroubleshootingBefore sending the pyrometer for repair, it is recommended that you try to find the erro

Page 19 - 4.16 Operating Hours

IS 320 • IGA 320 Manual Troubleshooting  26To ensure consistent document formatting, this page was intentionally left blank.

Page 20

IS 320 • IGA 320 Manual Data format UPP  278 Data format UPP (Universal PyrometerProtocol)You can communicate directly with the pyrometer through th

Page 21 - 5 Software InfraWin

IS 320 • IGA 320 Manual Data format UPP  28Reading basictemperature range:AAmbOutput: XXXXYYYY (hex 8-digit, °C or °F)XXXX = beginning of temp. rang

Page 22

IS 320 • IGA 320 Manual Data format UPP  29Detailed Softwareversion:AAvstt.mm.yy XX.YYtt = day; mm = month; yy = year;XX.YY = software versionRefere

Page 23 - 6 Maintenance

IS 320 • IGA 320 Manual Contents  iiiContents1 General ...

Page 25 - 7 Troubleshooting

IS 320 • IGA 320 Manual Reference Numbers  319 Reference Numbers9.1 Instrument Reference NumbersIS 320OpticsTemperature Range550–1400°C600–1600°C650

Page 26

IS 320 • IGA 320 Manual Reference Numbers  323 852 610 USB LabKit, adapter RS485 to USB with targeting light push-button and analogoutput clamp, py

Page 27 - Protocol)

IS 320 • IGA 320 Manual Index  33IndexAAccessories 10, 31Address 18Ambient temperature compensation 18Analog output 18Appropriate use 9BBasic Range

Page 28

IS 320 • IGA 320 Manual Index  34Subrange 18TTechnical data 9Technical Data 9Temperature display in °C or °F 15transmittance 16Transmittance  16Tra

Page 29

IS 320 • IGA 320 Manual Contents  iv4.7 EmiAutoFind ...

Page 30

IS 320 • IGA 320 Manual General  51 General1.1 Information about the user manualCongratulations on your LumaSense Technologies, Inc. IMPAC brand pyr

Page 31 - 9 Reference Numbers

IS 320 • IGA 320 Manual General  61.2.1 Electrical connectionFollow common safety regulations for mains voltage (230 or 115 V AC) connecting additio

Page 32

IS 320 • IGA 320 Manual General  7number. Upon receipt of this information, Technical Support will attempt to locate the faultand, if possible, solv

Page 33

IS 320 • IGA 320 Manual General  81.7 Transport, packaging, storageThe instrument can be damaged or destroyed if shipped incorrectly. To transport o

Page 34

IS 320 • IGA 320 Manual General  92 Introduction2.1 Appropriate useThe IS 320 and IGA 320 digital IMPAC pyrometers are designed for non-contact temp