OPERAT I ON MANUALIMPAC PyrometerIS 320 IGA 320
IS 320 • IGA 320 Manual General 10Repeatability:( =1, t90=1 s,Tamb.=23°C)0.2% of measured value in °C + 1 °CProtectionClass:IP65 (IEC 60529)Ambien
IS 320 • IGA 320 Manual General 113 Controls and Installation3.1 Electrical InstallationThe pyrometers are powered by a voltage of 24 V DC (possibl
IS 320 • IGA 320 Manual General 12Connecting to the RS485 Interface/BaudrateThe pyrometer is operated in half-duplex mode. To prevent reflectionsdu
IS 320 • IGA 320 Manual General 13becomes diffuse. The targeting light only marks the center of the measuring spot, not the size.A measuring tape c
IS 320 • IGA 320 Manual General 14The spot sizes in the table abovewill only be achieved at theadjusted distances. Decreasing orincreasing the meas
IS 320 • IGA 320 Manual General 154 Settings/Parameter DescriptionsThe pyrometer is equipped with a wide range of settings to ensure accurate tempe
IS 320 • IGA 320 Manual General 16same temperature. Different materials have different emissivities, ranging from 0% to 100%.The emissivity of an o
IS 320 • IGA 320 Manual General 174.9 Clear Time of the Maximum/Minimum Value Storage (tCL)If maximum value storage is on, the highest last tempera
IS 320 • IGA 320 Manual General 184.9.1 Max/MinWhen entering a reset time interval other than 0, it must also be determined with “max” or“min” if t
IS 320 • IGA 320 Manual General 19 Close above: The contact closes (switches to ground), if the temperature entered in the“SP1“ box is exceeded. I
Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us
IS 320 • IGA 320 Manual General 20 To ensure consistent document formatting, this page was intentionally left blank.
IS 320 • IGA 320 Manual Software InfraWin 215 Software InfraWinThe operating and analyzing InfraWin software is included with delivery of the pyrom
IS 320 • IGA 320 Manual Software InfraWin 22To ensure consistent document formatting, this page was intentionally left blank.
IS 320 • IGA 320 Manual Maintenance 236 Maintenance6.1 SafetyAttention during service to the pyrometer: If the pyrometer is integrated into arunnin
IS 320 • IGA 320 Manual Maintenance 24To ensure consistent document formatting, this page was intentionally left blank.
IS 320 • IGA 320 Manual Troubleshooting 257 TroubleshootingBefore sending the pyrometer for repair, it is recommended that you try to find the erro
IS 320 • IGA 320 Manual Troubleshooting 26To ensure consistent document formatting, this page was intentionally left blank.
IS 320 • IGA 320 Manual Data format UPP 278 Data format UPP (Universal PyrometerProtocol)You can communicate directly with the pyrometer through th
IS 320 • IGA 320 Manual Data format UPP 28Reading basictemperature range:AAmbOutput: XXXXYYYY (hex 8-digit, °C or °F)XXXX = beginning of temp. rang
IS 320 • IGA 320 Manual Data format UPP 29Detailed Softwareversion:AAvstt.mm.yy XX.YYtt = day; mm = month; yy = year;XX.YY = software versionRefere
IS 320 • IGA 320 Manual Contents iiiContents1 General ...
IS 320 • IGA 320 Manual Reference Numbers 319 Reference Numbers9.1 Instrument Reference NumbersIS 320OpticsTemperature Range550–1400°C600–1600°C650
IS 320 • IGA 320 Manual Reference Numbers 323 852 610 USB LabKit, adapter RS485 to USB with targeting light push-button and analogoutput clamp, py
IS 320 • IGA 320 Manual Index 33IndexAAccessories 10, 31Address 18Ambient temperature compensation 18Analog output 18Appropriate use 9BBasic Range
IS 320 • IGA 320 Manual Index 34Subrange 18TTechnical data 9Technical Data 9Temperature display in °C or °F 15transmittance 16Transmittance 16Tra
IS 320 • IGA 320 Manual Contents iv4.7 EmiAutoFind ...
IS 320 • IGA 320 Manual General 51 General1.1 Information about the user manualCongratulations on your LumaSense Technologies, Inc. IMPAC brand pyr
IS 320 • IGA 320 Manual General 61.2.1 Electrical connectionFollow common safety regulations for mains voltage (230 or 115 V AC) connecting additio
IS 320 • IGA 320 Manual General 7number. Upon receipt of this information, Technical Support will attempt to locate the faultand, if possible, solv
IS 320 • IGA 320 Manual General 81.7 Transport, packaging, storageThe instrument can be damaged or destroyed if shipped incorrectly. To transport o
IS 320 • IGA 320 Manual General 92 Introduction2.1 Appropriate useThe IS 320 and IGA 320 digital IMPAC pyrometers are designed for non-contact temp