MANUA LM330Blackbody Calibration Source
M330 Manual Introduction 102.4 Dimensions330M330 Front Panel
M330 Manual Introduction 11M330 Back Panel
M330 Manual Introduction 12To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Installation 133 InstallationInstall the blackbody emitter in the desired location that preferably has no moving air near theemitter ap
M330 Manual Installation 145. As the cavity flange approaches the furnace tube, very gently continue insertion untilthe cavity stops moving. It is
M330 Manual Operation 154 Operation4.1 Starting Up1. Allow the M330 to warm-up to room temperature for at least two hours afterunpacking and prior
M330 Manual Operation 164.2 Changing the SetpointPress either the UP/DWN buttons on the controller. Each press will advance the setpoint 1degree. T
M330 Manual Operation 174.4 Changing Temperature UnitsYou can select one of three temperature units: Celsius, Fahrenheit, and Kelvin.To change temp
M330 Manual Operation 184.5.2 RS485 OptionRS485 two wire is available as an option. The connection scheme is shown below:RS485 for multi-drop appl
M330 Manual Operation 194.6 PID Tuning ConstantsThe PID tuning constants have been factory set for optimum performance and are noted below.PID 1 PI
Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us
M330 Manual Operation 20To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Maintenance 215 Maintenance5.1 Cavity LifeThe life expectancy of the cavity depends on the time and temperature of operation. The highe
M330 Manual Maintenance 22To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Service 236 Service6.1 Controller Calibration (Adjustment)The M330 has a precision, thermocouplemonitoring the emitter temperature. Th
M330 Manual Service 242. Press 2.3. Press 4 until Level 3 appears.4. Press 4 until 3 appears. Wait.5. Press 1 until INPUT OPERS and Cust Lin 1 Page
M330 Manual Service 250-510°CPID1(PB1, Int1, Rate1)510-710°CPID2(PB2, Int2, Rate2)710-1380°CPID3(PB3, Int3, Rate3)1380-1750°C(end)PID4(PB4, Int4, R
M330 Manual Service 266.3.1 Primary Thermocouple Replacement Procedure1. Remove the screws holding the M330 front and rear panels.2. Lower the fron
M330 Manual Service 27Warning: Electrical Shock Hazard! This equipment is to be serviced by trainedpersonnel only.8. Replace with new thermocouple
M330 Manual Service 28To check which heating element is broken:1. Disconnect the AC power.2. Locate which heating element is broken. Disconnect on
M330 Manual Service 29To replace a heating element:To replace a heater, both thermocouples must be moved out of the way of the heaters. Thenthe to
M330 Manual Contents iiiContents1 General Information ...
M330 Manual Service 305. Pull top cover with heaters AND the three insulation blocks up and out.6. Remove only the bad heating element by loosening
M330 Manual Service 3110. Replace straps.11. Replace main supply wires to end straps.Proper connection of electrical straps and clamps. Top view.
M330 Manual Service 326.5 Outer Tube ReplacementWarning: Electrical Shock Hazard! This equipment is to be serviced by trainedpersonnel only.1. Loos
M330 Manual Service 334. Remove (chip away) ceramic cement at this location with small hammer and screwdriver.5. Slide outer tube out of cavity. Re
M330 Manual Service 346.6 Field Modification for Low Heat RateWarning: Electrical Shock Hazard! This equipment is to be serviced by trainedpersonne
M330 Manual Service 35Top view of new Thermostat location.Rear of blackbodyFront of blackbody3. Move Transformer Primary tap from 230 VAC (H3) to 2
M330 Manual Service 36Warning: Electrical Shock Hazard! This equipment is to be serviced by trainedpersonnel only.4. Program M330 Controller accord
M330 Manual Service 371. Remove rear cover.2. Attach a Clamp-On Ammeter to one of the supply wires. This meter should measure TrueRMS current. Typi
M330 Manual Service 38To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Index 39IndexAAppropriate use 9BBaud Rate 17CCavity life 21Cleaning 21Controller Adjustment 23Controller Calibration 23DDimensions 10Di
M330 Manual Contents iv6.3.2 Secondary Thermocouple Replacement Procedure... 276.4 Heater Element Replaceme
M330 Manual Index 40To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Appendix A: M330 Spectral Correction Procedure 41Appendix A: M330 Spectral CorrectionProcedure1. Set the M330 controller to the desired
M330 Manual Appendix A: M330 Spectral Correction Procedure 42950 948.1 943.2 940.9 937.3 933.9 935.4 935.7 942.1975 973.1 967.9 965.6 961.8 958.3 9
M330 Manual Appendix B: Optional Blackbody Spectral Radiance Correction Software 43Appendix B: Optional Blackbody SpectralRadiance Correction Soft
M330 Manual Appendix B: Optional Blackbody Spectral Radiance Correction Software 44To ensure consistent document formatting, this page was intenti
M330 Manual Appendix C: CE Certification 45Appendix C: CE Certification
M330 Manual General Information 51 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly
M330 Manual General Information 6 Do not attempt to operate with any of the safety features bypassed or disconnected. Do not place this unit wher
M330 Manual General Information 7Santa Clara, California Telephone: +1 408 727 1600 or +1 800 631 0176 Email: [email protected],
M330 Manual General Information 8To ensure consistent document formatting, this page was intentionally left blank.
M330 Manual Introduction 92 Introduction2.1 DescriptionThe M330 is a blackbody calibration source utilizing a digital indicating temperature contro