Lumasense-technologies M310 User Manual

Browse online or download User Manual for Tools Lumasense-technologies M310. LumaSense Technologies M310 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Blackbody Calibration Source

MANUA LM310Blackbody Calibration Source

Page 2

M310 Manual Introduction  102.4 DimensionsM310 Front PanelM310 Back Panel

Page 3 - Contents

M310 Manual Installation  113 Installation3.1 LocationSelect a stable location on which to place the M310. Since any blackbody source stability can

Page 4 - M310 Manual Contents  iv

M310 Manual Installation  12To ensure consistent document formatting, this page was intentionally left blank.

Page 5 - 1 General Information

M310 Manual Operation  134 Operation4.1 Starting Up1. Allow the M310 to warm-up to room temperature for at least two hours afterunpacking and prior

Page 6 - 1.4 Unpacking the Instrument

M310 Manual Operation  144.4 Changing Temperature Units (C-F, F-C)The controller has been locked out except for setpoint change to hinder tampering

Page 7

M310 Manual Operation  158. Press UP to select according to chart below.9. Press SCROLL and hold for 3 seconds to exit.Note: It is important to loc

Page 8

M310 Manual Operation  16To ensure consistent document formatting, this page was intentionally left blank.

Page 9 - 2 Introduction

M310 Manual Communications Port  175 Communications Port5.1 RS232CThis blackbody may have the option to communicate with the temperature controller

Page 10 - 2.4 Dimensions

M310 Manual Communications Port  18RS485 wiringThe blackbody was shipped with the controller set at the following com port parameters: Baud: 1200

Page 11 - 3 Installation

M310 Manual Communications Port  19Controller response:%0101R0902.000N8<CR>Command line to change blackbody setpoint to 30C:$0101W0930.000F5&l

Page 12

Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us

Page 13 - 4 Operation

M310 Manual Communications Port  20To ensure consistent document formatting, this page was intentionally left blank.

Page 14 - C to F or F to C

M310 Manual Maintenance  216 Maintenance6.1 Cavity LifeThe life expectancy of the cavity depends on the time and temperature of operation. The highe

Page 15 - 4.6 PID Tuning Constants

M310 Manual Maintenance  22To ensure consistent document formatting, this page was intentionally left blank.

Page 16 - M310 Manual Operation  16

M310 Manual Index  23IndexAAir Outlet Clearance 11Apparent Emissivity 9Appropriate use 9BBackground Influence 9Baud Rate 14CCavity life 21Cleaning 2

Page 17 - 5 Communications Port

M310 Manual Index  24To ensure consistent document formatting, this page was intentionally left blank.

Page 18 - 5.3 Communication Command Set

M310 Manual Appendix A: Sample Communication Program  25Appendix A: Sample CommunicationProgramThis is a typical QBASIC program that reads current t

Page 19

M310 Manual Appendix A: Sample Communication Program  26 PRINT USING "##.###"; getpval LOCATE 6, 40 PRINT "new setpoint in&q

Page 20

M310 Manual Appendix B: Calibration Methods  27Appendix B: Calibration MethodsThere are two distinctly different methods of calibration, each with i

Page 21 - 6 Maintenance

M310 Manual Appendix B: Calibration Methods  28Emissivity (for M310, M315, M320, M316 series blackbodies)When measurements are required at wavelengt

Page 22 - M310 Manual Maintenance  22

M310 Manual Appendix B: Calibration Methods  29When calibrating infrared sensors at wavelengths other than 8-14 µm region, the graph belowdisplays t

Page 23 - M310 Manual Index  23

M310 Manual Contents  iiiContents1 General Information ...

Page 24 - M310 Manual Index  24

M310 Manual Appendix B: Calibration Methods  30Example 1: The HT (450 C) blackbody was factory calibrated to provide effective emissivity =1.0 at 8

Page 25

M310 Manual Appendix B: Calibration Methods  31Example 2: The standard (350 C) blackbody was factory calibrated to provide effectiveemissivity = 1.

Page 26

M310 Manual Appendix B: Calibration Methods  32Example 3: The HT (450 C) blackbody was factory calibrated to provide effective emissivity =1.0 at 8

Page 27

M310 Manual Appendix B: Calibration Methods  33Using the M310 to calibrate “fixed emissivity” pyrometersFor convenience, the table below shows what

Page 28

M310 Manual Appendix B: Calibration Methods  34To ensure consistent document formatting, this page was intentionally left blank.

Page 29

M310 Manual Appendix C: CE Certification  35Appendix C: CE Certification

Page 30

M310 Manual Contents  iv6.6 Heater Element Replacement ... 21Index ...

Page 31

M310 Manual General Information  51 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly

Page 32

M310 Manual General Information  61.3 Limit of Liability and WarrantyAll general information and notes for handling, maintaining, and cleaning this

Page 33

M310 Manual General Information  7 Telephone +33 (0)3 88 98 98 01 Email: [email protected] Shipments to LumaSense for RepairAll RMA sh

Page 34

M310 Manual General Information  8To ensure consistent document formatting, this page was intentionally left blank.

Page 35 - Appendix C: CE Certification

M310 Manual Introduction  92 Introduction2.1 DescriptionThe M310 is a portable blackbody calibration source utilizing a digital indicating temperatu