MANUA LIMPAC PyrometersISR 6 Advanced

Page 2

ISR 6 Advanced Manual Introduction · 10OpticsSighting: Built-in laser aiming light (max. power level < 1 mW,l = 630-680 nm, CDRH class II) or thr

Page 3 - Contents

ISR 6 Advanced Manual Introduction · 11PowerConsumption:Max. 3 W ( incl. laser)Load(analog output):0 to 500 W Isolation:Power supply, analog output a

Page 4

ISR 6 Advanced Manual Introduction · 122.6 Accessories (optional)Numerous accessories guarantee easy installation of the pyrometer. The following ove

Page 5 - 1 General Information

ISR 6 Advanced Manual Introduction · 13It can consist of e.g. mounting support, tube support with air purge and flange and an open orclosed ceramic s

Page 6 - 1.4 Unpacking the Instrument

ISR 6 Advanced Manual Introduction · 14To ensure consistent document formatting, this page was intentionally left blank.

Page 7

ISR 6 Advanced Manual Controls and Installation · 153 Controls and Installation3.1 Electrical InstallationThe pyrometer is powered by 24 V DC ± 25% (

Page 8

ISR 6 Advanced Manual Controls and Installation · 16The connector pin J can be used for 3 different functions:1. External clearing of the maximum val

Page 9 - 2 Introduction

ISR 6 Advanced Manual Controls and Installation · 173.1.3 Connection to RS485Half-duplex mode: A1 and A2 as well as B1 and B2 are bridged in the 12-p

Page 10

ISR 6 Advanced Manual Controls and Installation · 18Instruments like an analog controller or printer can be connected to the display in a series assh

Page 11 - 2.5 Physical User Interface

ISR 6 Advanced Manual Controls and Installation · 19Note: The laser warning signs on the pyrometer should be easily viewable at all times,even after

Page 12 - 2.6 Accessories (optional)

Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us

Page 13

ISR 6 Advanced Manual Controls and Installation · 20Focused spot sizes between the listed distances can be found by linear interpolation betweenthe l

Page 14

ISR 6 Advanced Manual Settings / parameter descriptions · 214 Settings / parameter descriptionsThe pyrometer is equipped with a wide range of setting

Page 15 - 3 Controls and Installation

ISR 6 Advanced Manual Settings / parameter descriptions · 22using a thermocouple probe or by knowing one process temperature point precisely from oth

Page 16

ISR 6 Advanced Manual Settings / parameter descriptions · 23The table shows typical errors that can result when the emissivity of one wavelength diff

Page 17 - 3.1.3 Connection to RS485

ISR 6 Advanced Manual Settings / parameter descriptions · 24If the maximum value storage is switched on, the highest last temperature value will alwa

Page 18 - 3.2 Sighting

ISR 6 Advanced Manual Settings / parameter descriptions · 25AUTO The AUTO mode is used for discontinuous measuring tasks. For example,when objects ar

Page 19 - 3.3 Optics

ISR 6 Advanced Manual Settings / parameter descriptions · 26Note: Settings for Minimum Intensity Switch-Off Level: 2% to 50%Ratio pyrometers are abl

Page 20

ISR 6 Advanced Manual Settings / parameter descriptions · 27you to program all pyrometers at the same time, regardless of the addresses that have alr

Page 21 - 4.3 Emissivity Slope K

ISR 6 Advanced Manual Settings / parameter descriptions · 28To ensure consistent document formatting, this page was intentionally left blank.

Page 22 - 4.3.1 Slope Adjustment

ISR 6 Advanced Manual Software InfraWin · 295 Software InfraWinThe operating and analyzing InfraWin software is included with delivery of the pyromet

Page 23 - 4.6 Clear Peak Memory (t

ISR 6 Advanced Manual Contents · iiiContents1 General Information ...

Page 24 - 4.6.2 Clear Time Settings

ISR 6 Advanced Manual Software InfraWin · 305.4 Basic settingsAll preset values for the devicecan be displayed and modified,if necessary under theDev

Page 25 - 4.9 “Dirty Window” Warning

ISR 6 Advanced Manual Software InfraWin · 31In the lower part of the window, the connection with the preset baud rate can be checked. Herethe command

Page 26 - 4.13 Device Address

ISR 6 Advanced Manual Software InfraWin · 325.6 Warning Level / Switch-Off LevelIn addition to the relative signal bar graph, the warning level and t

Page 27 - 4.15 Baud Rate

ISR 6 Advanced Manual Software InfraWin · 33Once the temperature entry has been enteredand confirmed with OK, InfraWin will thencalculate the emissiv

Page 28

ISR 6 Advanced Manual Software InfraWin · 34the amount of data increases, so does the amount of storage space required to save it. In orderto save ro

Page 29 - 5 Software InfraWin

ISR 6 Advanced Manual Software InfraWin · 355.14 Search I/O ModuleThe I/O Module allowsaccessories to connect tothe software and is used to triggerme

Page 30 - 5.4 Basic settings

ISR 6 Advanced Manual Software InfraWin · 36To ensure consistent document formatting, this page was intentionally left blank.

Page 31 - 5.5 Relative Signal

ISR 6 Advanced Manual Data format UPPÒ· 376 Maintenance6.1 Cleaning ISR 6 WindowBecause there are no moving parts in the ISR 6, the only regular main

Page 32

ISR 6 Advanced Manual Data format UPPÒ· 38In the 1-color (mono) mode, the pyrometer can measure objects at any distance.However, the object has to be

Page 33 - 5.9 Listing (analyzing)

ISR 6 Advanced Manual Data format UPPÒ· 397 Data format UPPÒ (Universal PyrometerProtocol)Software commands can be exchanged directly with the pyrome

Page 34 - 5.13 Spot size calculator

ISR 6 Advanced Manual Contents · iv4.4 Transmittance t...

Page 35 - 5.14 Search I/O Module

ISR 6 Advanced Manual Data format UPPÒ· 40Description Command ParametersMeasuring value(one-channel andratio temperature)AAek Answer: SSSSSQQQQQ2x5 d

Page 36

ISR 6 Advanced Manual Data format UPPÒ· 41Description Command ParametersRead parameters AApa Answer: 15 decimal digitsDD . . . . . . . . . . . . .

Page 37 - 6 Maintenance

ISR 6 Advanced Manual Data format UPPÒ· 42To ensure consistent document formatting, this page was intentionally left blank.

Page 38

ISR 6 Advanced Manual Reference Numbers · 438 Reference Numbers8.1 Reference numbers instrumentTemperature Range With Viewfinder With Laser Targetin

Page 39 - Protocol)

ISR 6 Advanced Manual Reference Numbers · 443 852 550Power supply NG 2D for DIN rail mounting; 85 to 265 V AC Þ 24 V DC, 600 mAwith two settable limi

Page 40

ISR 6 Advanced Manual Troubleshooting · 459 TroubleshootingSymptom Probable Cause CommentsNo analog output evenif the display shows atemperature abov

Page 41

ISR 6 Advanced Manual Troubleshooting · 46Symptom Probable Cause CommentsTemperature readingsare too highTargets hotter than expected.(This often hap

Page 42

ISR 6 Advanced Manual Troubleshooting · 47Symptom Probable Cause CommentsNoisy readings (fastfluctuations).Target moves more than 95%out of field of

Page 43 - 8 Reference Numbers

ISR 6 Advanced Manual Troubleshooting · 48To ensure consistent document formatting, this page was intentionally left blank.

Page 44

ISR 6 Advanced Manual Index · 49Index11 measure… 30AAccesories 12Air Purge 12Analog Output 25Appropriate use 9BBasic settings 30Baud Rate 27CCalibrat

Page 45 - 9 Troubleshooting

ISR 6 Advanced Manual General Information · 51 General Information1.1 Information about the user manualCongratulations on choosing the high quality a

Page 46

ISR 6 Advanced Manual · 50Response Time 23RS485 17SSafety 5Scaling Trend button 33Scope of delivery 9Service Request 6Settings / parameter descriptio

Page 47

ISR 6 Advanced Manual General Information · 6The warranty is VOID if the instrument is disassembled, tampered with, altered, or otherwisedamaged with

Page 48

ISR 6 Advanced Manual General Information · 7Technical Support can be contacted by telephone or email:Santa Clara, California· Telephone: +1 408 727

Page 49

ISR 6 Advanced Manual General Information · 8To ensure consistent document formatting, this page was intentionally left blank.

Page 50 - ISR 6 Advanced Manual · 50

ISR 6 Advanced Manual Introduction · 92 Introduction2.1 Appropriate useThe IMPAC ISR 6 pyrometer is a stationary, digital pyrometer for non-contact t

Comments to this Manuals

No comments

VADDIO WideSHOT WallVIEW USB manuals

Owner’s manuals and user’s guides for Audio VADDIO WideSHOT WallVIEW USB.
We providing 1 pdf manuals VADDIO WideSHOT WallVIEW USB for download free by document types: User Manual






More products and manuals for Audio VADDIO

Models Document Type
ZoomSHOT WallVIEW Mini User Manual   Vaddio ZoomSHOT WallVIEW Mini User Manual, 32 pages
ZoomSHOT WallVIEW SR User Manual   Vaddio ZoomSHOT WallVIEW SR User Manual, 16 pages
ZoomSHOT WallVIEW USB User Manual   Vaddio ZoomSHOT WallVIEW USB User Manual, 40 pages